Max-11 Oxygen Sensor

Replacement For

  • Analytical Industries [PSR 11 915 2]
  • Drager [6803290, 6850645, Narkomed G5, Fabius GS, Evita Family, Babylog, Cato, Fabius Series]
  • EnviteC [OOM-201]
  • HillRom [PSR-11-915-2]
  • International Technologies [M-02]
  • Invivo [Magnitude]
  • Mercury Medical [10-103-07]
  • Pacifitech [PT-14A]
  • Teledyne [R23 #C-61139]
  • Teleflex (Hudson) [5557, V-14]
  • VTI [101372, 100352, 101372, 100352]

Specifications

  • Range of Measurement: 0 to 100%
  • Output Range: 14 to 20 mV
  • Zero Offset: ≤ 0.20 mV
  • Response Time: ≤ 12 s
  • Linearity: ≤ 3%
  • Expected Lifetime: > 500,000 %O2 hours
  • Temperature Compensation: none
  • Operational Temperature Range: 10˚ to 40˚C
  • Storage Temperature Range: -20˚ to 50˚C
  • Humidity Range: 0% to 99% RH, non-condensing
  • Ambient Pressure Range: 700-1250 mbar
  • Interference: PER EN ISO 80601-2-55
  • Electrical Interface: gold-plated slip ring