Max-18

Replacement For

  • Analytical Industries [PSR-11-33-1]
  • Mercury Medical [10-103-01]
  • Pacifitech [PT-02]
  • Teledyne [C1R #B-39022]
  • Teleflex (Hudson) [5569, V-02, 5525, 5528, 5529, 5550, 5552]

Specs

  • Range of Measurement: 0 to 100% O2
  • Output (in air): 200 ± 40 µA
  • Zero Offset: Less than 5 µA
  • Response Time: ≤ 13 seconds at 30° C
  • Linearity: < 3%
  • Expected Lifetime: > 250,000 %O2 hours
  • Operational Temperature Range: 0° to 40° C
  • Storage Temperature Range: -20° to 50° C
  • Humidity Range: Up to 100% RH
  • Ambient Pressure Range: 600 - 1750 mBar
  • Interference: Per EN ISO 80601-2-55
  • Electrical Interface: 1mm x 3.5mm Pins